Improve OCT Image Quality on Monaco and Silverstone

The Signal to Noise Ratio (SNR) value has a great impact on the quality of your Monaco OCT images. A high SNR will display a full and dense scan, while a low SNR will result in a grainy image. SNR should be at a minimum of 6/10. To improve the SNR, press the plus (+) or minus (-) button on your hand controller.

In cases where the location, OCT scan and/or SNR need readjustment, select OCT setup above the right side of the OCT scan. You have the option of re-optimizing the location, focus, or OCT individually or re-optimizing all. Please access your device Instructions for Use (IFU) for full device use instructions. IFUs can be found in CustomerCentral

Note: Please log in CustomerCentral or create an account to download the instructions below.
Download the quick reference guide for the Monaco device.
Download the quick reference guide for the Silverstone device.

 